Mankar, Swapnali Suresh and Younas, Muhammad and Kalode, Vaishavi and Jirvankar, Pranita and Phutane, Prasanna and Chandewar, Shital and Armarkar, Ashwini and Pethe, Anil (2021) Selective SGLT2 Inhibitor’s Estimation and Validation from Galenical Form by RP-HPLC Method. Journal of Pharmaceutical Research International, 33 (57B). pp. 57-68. ISSN 2456-9119

[thumbnail of 4811-Article Text-6911-1-10-20221006.pdf] Text
4811-Article Text-6911-1-10-20221006.pdf - Published Version

Download (517kB)

Abstract

Aim: The recent analysis is required to do novel and simple, sensitive, precise, efficient, instant and reproducible reverse phase high performance liquid chromatography (RP-HPLC) method for estimation of antidiabetic drug in the unit dosage form. Validation of this method is also planned to make it suitable for the actual use.

Study Design: Department of Pharmaceutical Chemistry, Datta Meghe Institute of Medical Sciences deemed to be university, Wardha in collaboration with Balkh university, Mazar-e-sharif, Afghanistan between August 2021 and December 2021.

Methodology: In that article develop the method and validate it by estimation of antidiabetic drugs in solid dosage form by RP-HPLC, by using System suitability test, Repeatability, Precision studies (Intra-day and Interday/Intermediate), Linearity/Calibration studies, Robustness, Force degradation, Specificity, Drug recovery/accuracy studies.

Results: as per ICH guidelines, the performance if system suitability in remogliflozin were achieved all guidelines; in that, tailing factor (T),separation factors(α),theoretical plates(N),capacity factor (k’), resolution (R) and RSD (%). The validated stress degradation studies under thermal, oxidative, alkali and acid, in few degradation products for remogliflozin (REM).

Conclusion: From the results we conclude that, this novel technique which validated for exploration by reverse phase high performance liquid chromatography (RP-HPLC) should be used for routine quality control of remogliflozin (REM) prediction from developed formulation.